Indian Journal of Dental Research

ORIGINAL RESEARCH
Year
: 2013  |  Volume : 24  |  Issue : 5  |  Page : 542--546

Evaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensor


Denise Pupim1, Liogi Iwaki Filho2, Wilton Mitsunari Takeshita3, Lilian Cristina Vessoni Iwaki2 
1 Department of Dentistry, University of São Paulo, School of Dentistry of Ribeirão Preto, Ribeirão Preto, Brazil
2 Department of Dentistry, State University of Maringá, Maringá, Brazil
3 Department of Dentistry, Federal University of Sergipe, Aracaju, Brazil

Correspondence Address:
Wilton Mitsunari Takeshita
Department of Dentistry, Federal University of Sergipe, Aracaju
Brazil

Context: The pulse oximeter is a device that noninvasively provides continuous information about the peripheral oxygen saturation (SpO2) rate. This device is utilized in the detection of hypoxemia, due to its able to sense changes in hemoglobin oxygen saturation. Aims: The objective of this study was to verify the accuracy of the Choice® Medical MD300C3 Fingertip Pulse Oximeter, as compared to that of a hospital oximeter coupled with a Dräger® Infinity Delta monitor, with the purpose of using this first methodology in dental procedures to monitor the peripheral oxygen saturation (SpO2) of patients submitted to dental treatments. Materials and Methods: Fifty-five adult patients, both genders, were selected in the Santa Casa Hospital of Maringa, Brazil. The volunteers did not present cardiac problems, prosthetic cardiac valves, pacemakers, or pulmonary diseases, and were not pregnant or children. Each patient received a portable fingertip pulse oximeter (PPO) on the middle finger of the left hand and the hospital oximeter (control device) on the forefinger of the same hand. A total of six measurements were developed. The Pearson correlation coefficient and the Bland and Altman method was used to calculate the statistical analysis. Results: No statistically significant difference was found between the measurements taken by the utilized devices. The average of comparative analysis presented by the devices was 0.2337 ± 0.4355 (mean ± SD), suggesting a strong correlation between the obtained results. Conclusion: According to the methodology of the research, the PPO has similar accuracy to the conventional hospital oximeter with digital sensor. The PPO can be used in dental treatments.


How to cite this article:
Pupim D, Filho LI, Takeshita WM, Iwaki LV. Evaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensor.Indian J Dent Res 2013;24:542-546


How to cite this URL:
Pupim D, Filho LI, Takeshita WM, Iwaki LV. Evaluation of accuracy of portable fingertip pulse oximeter, as compared to that of a hospital oximeter with digital sensor. Indian J Dent Res [serial online] 2013 [cited 2020 Oct 22 ];24:542-546