Indian Journal of Dental Research

ORIGINAL RESEARCH
Year
: 2011  |  Volume : 22  |  Issue : 2  |  Page : 362-

Reliability of Logicon caries detector in the detection and depth assessment of dental caries: An in-vitro study


Rohit R Behere, Shailesh M Lele 
 Department of Oral Medicine and Radiology, Sinhgad Dental College and Hospital, Pune, India

Correspondence Address:
Rohit R Behere
Department of Oral Medicine and Radiology, Sinhgad Dental College and Hospital, Pune
India

Background: Digital radiography has so far not resulted in improved rates of proximal caries detection. Historically, automated caries detection tools have been largely academic. Opinions regarding the performance of the only such commercially available tool, viz., Logicon caries Detector (LCD) have been equivocal. This study was conducted to evaluate the reliability of LCD in the detection and depth assessment of proximal caries. Materials and Methods: Digital images were obtained of 100 proximal tooth surfaces using the Kodak RVG 5000 sensor and analyzed by three observers. The images were then analyzed by the principal investigator using the LCD software. The teeth were then sectioned and magnified photographic images were obtained which were taken as the gold standard. All the grades were entered in proformas and the data were statistically analyzed using the chi-square test. Five parameters of reliability were calculated. Results: The sensitivity, specificity, positive predictive value, negative predictive value, and overall accuracy of LCD for the grade No caries were 33, 96, 73, 82, and 81%, respectively; for the grade Enamel caries were 5, 97, 33, 80, and 79%, respectively; and for the grade Dentin caries were 100, 96, 50, 100, and 96%, respectively. Conclusions: In conclusion, LCD appears to be more reliable in ruling out (both enamel and dentin) caries than in detecting caries.


How to cite this article:
Behere RR, Lele SM. Reliability of Logicon caries detector in the detection and depth assessment of dental caries: An in-vitro study.Indian J Dent Res 2011;22:362-362


How to cite this URL:
Behere RR, Lele SM. Reliability of Logicon caries detector in the detection and depth assessment of dental caries: An in-vitro study. Indian J Dent Res [serial online] 2011 [cited 2020 Aug 13 ];22:362-362
Available from: http://www.ijdr.in/article.asp?issn=0970-9290;year=2011;volume=22;issue=2;spage=362;epage=362;aulast=Behere;type=0