ORIGINAL RESEARCH  
Year : 2018  |  Volume : 29  |  Issue : 5  |  Page : 634-640
A three-dimensional finite element analysis to evaluate stress distribution tooth in tooth implant-supported prosthesis with variations in non-rigid connector design and location


1 Department of Prosthodontics and Crown and Bridge, Hitkarini Dental College and Hospital, Jabalpur, Madhya Pradesh, India
2 Former Professor and HOD, Faculty of Dental Sciences, Sri Ramachandra University, Chennai, Tamil Nadu, India
3 Department of Prosthodontics and Crown and Bridge, Faculty of Dental Sciences, Sri Ramachandra University, Chennai, Tamil Nadu, India

Click here for correspondence address and email

Date of Web Publication2-Nov-2018
 

   Abstract 

Background: The physiologically mobile natural tooth and rigidly fixed dental implant causes different distribution of stress when connected in prosthesis and nonrigid connector compensates this. Understanding of biomechanical behavior is necessary for an adequate choice and construction of this type of rehabilitation. However, there has been insufficient research focusing on different location and type of the nonrigid connector related with the prognosis of both implant and the tooth. Aim of the Study: The purpose of this finite element (FE) analysis was to evaluate the stress distribution around bone, implant, and tooth in tooth implant fixed prosthesis under static load with variations in design and location of nonrigid connectors under simulated functional loads. Materials and Method: Three, 3-dimensional FE models connecting tooth and implant were constructed with different location and type of nonrigid connector. Simulated occlusal load was applied on the restorations and stresses developed in the supporting structures were monitored. Results: The highest stresses were found around the implant in model with nonrigid connector placed between the tooth and implant and model with modified nonrigid connector. On the other hand, less stress was noted around the implant where nonrigid connector was placed between the implant and pontic. Conclusion: It is advisable to place the nonrigid connector between the implant and the pontic to protect the implant from torque effects in a tooth implant fixed prosthesis.

Keywords: Finite element analysis, nonrigid connector, tooth implant-supported prosthesis

How to cite this article:
Nitin K S, Padmanabhan T V, Kumar V A, Parthasarathi N, Uma Maheswari M, Kumar S M. A three-dimensional finite element analysis to evaluate stress distribution tooth in tooth implant-supported prosthesis with variations in non-rigid connector design and location. Indian J Dent Res 2018;29:634-40

How to cite this URL:
Nitin K S, Padmanabhan T V, Kumar V A, Parthasarathi N, Uma Maheswari M, Kumar S M. A three-dimensional finite element analysis to evaluate stress distribution tooth in tooth implant-supported prosthesis with variations in non-rigid connector design and location. Indian J Dent Res [serial online] 2018 [cited 2018 Dec 18];29:634-40. Available from: http://www.ijdr.in/text.asp?2018/29/5/634/244944

   Introduction Top


Implant dentistry is a revolutionary improvement in functional and esthetic rehabilitation of both partially and completely edentulous conditions. Although many treatment options such as cast partial denture and freestanding implant-supported restoration are available for distal extension situations, tooth implant fixed prosthesis has shown efficient treatment alternative for patients with anatomic and economic limitations.[1],[2] Teeth with a sound periodontal ligament have mobility characteristics between 50 and 200 μm, while osseointegrated implants demonstrate a mobility of <10 μm.[3],[4],[5],[6],[7] This difference in mobility has been associated with a number of problems such as intrusion of abutment teeth,[1],[8],[9] abutment screw loosening,[7],[8] fracture of abutment teeth, and fracture of implant components.[7],[10] This can be compensated by nonrigid connectors, which act as stress breaker with the ability to separate the splinted units. Cohen and Orenstein claims that the nonrigid connection between teeth and implant limits cantilever forces and directs occlusal loads axially to the implant.[11]

The different nonrigid connectors systems available are (1) implant with a stress-absorbing element (intramobile element or stress-breaking element)[12] and (2) Key and Keyway stress breakers which are commonly used with conventional FPDs.[13],[14],[15] Chee and Jivraj designed a nonrigid connection between implant and pontic with spoon-shaped rest from distal attachment to the pontic seated on tooth implant-supported restoration.[16],[17]

Theoretically, the suggested nonrigid connections between implant-supported crown and the pontic will transfer favorable occlusal load. However, there has been insufficient research focusing on the connector designs used on tooth implant fixed prosthesis. One of the important reasons is that many clinical observations could not have sufficient enough information to determine the biomechanics of complicated tooth implant supported systems. Computer simulations based on the finite element (FE) method has been employed as a tool for evaluating mechanical responses such as internal stresses or relative micromotions. Hence, three-dimensional (3D) FE analysis was used to assess the stress distribution around bone, natural tooth, and implant under occlusal load on tooth implant fixed prosthesis with different nonrigid connector design and location.


   Materials and Methods Top


A 3D FE analysis was conducted using mathematical models of posterior 3 unit tooth implant-supported prosthesis with a nonrigid connector. The different design and location of nonrigid connector were divided into 3 models as:

  • Model 1 - Nonrigid connector placed on mesial side of implant [Figure 1]
  • Model 2 - Nonrigid connector placed on distal side of premolar [Figure 2]
  • Model 3 - A passive connection in the form of a removable partial denture type spoon-shaped rest extending from the distal surface of the pontic and rest seat preparation on the restored implant. On the mesial side of the pontic, it was rigidly connected to the tooth [Figure 3].
Figure 1: Model 1 showing the nonrigid connector placed on mesial side of implant

Click here to view
Figure 2: Model 2 showing the nonrigid connector placed on distal side of premolar

Click here to view
Figure 3: Model 3 modified nonrigid connector as occlusal rest extending from the distal surface of the pontic and rest preparation on the restored implant and the mesial side of the pontic was rigidly connected to the tooth

Click here to view


A computed tomography image of adult mandible was used to make the 3D model of second premolar tooth with edentulous mandible distal to it. Modeling was done by using a Pro/Engineer Software (Parametric Corporation, Technology Needham, MA, USA) and exported to an analysis package. A bone block with height of 29 mm, width of 12 mm, and cortical bone thickness of 1.5 mm surrounding the cancellous bone was modeled. The model of premolar tooth was simulated with root length of 13 mm, crown length of 8 mm, mesiodistal and buccolingual diameter of 8 mm in accordance to Wheelers  Atlas More Details of Tooth form.[18] The periodontal ligament of 0.2 mm width was simulated. General rule of tooth preparation was followed for metal-ceramic restoration with chamfer margin.[19] Physical models of implant of length 13.5 mm and width 3.45 mm were converted to their geometric model and simulated to be placed in section of mandibular bone at the second molar region. The implants were directly in contact with the bone while natural tooth had mobility within the borders of the periodontal membrane.

A metal-ceramic restoration was simulated with Ni-Cr alloy substructure and finished with feldspathic porcelain for tooth implant-supported prosthesis. Two types of nonrigid connectors were used in the study - Beyeler intracoronal, friction grip, dovetail slide key and keyway attachment (Cendres and Metaux, Switzerland)[20] with the fixed length 3.8 mm. Other model of nonrigid connector was made with an occlusal rest extending from the distal surface of the pontic and a spoon-shaped rest preparation on the restored implant. The dimensions of rest seat were taken from the literature.[21] Both the non rigid connectors were allowed to move vertically on each other where their surfaces were in contact.

The cortical bone, the trabecular bone, implant and its abutment, and mucosa were modeled. The elastic modulus and Poisson's ratio of titanium, trabecular, cortical bones, periodontal ligament and dentin, and feldspathic porcelain were obtained from the literature[22] and listed in [Table 1]. All the materials used in the models were considered to be isotropic, homogeneous, and linearly elastic, and the osseointegration of the implants was accepted as 100%. The finite element software, ANSYS Workbench Software (Santa Monica, CA, USA) was used to analyze the models.
Table 1: Modulus of elasticity and Poisson's ratio of various materials used in the study

Click here to view


Boundary

Considering the support of masseter and medial pterygoid muscles that attach to the external angle of mandible, the boundary condition with no displacement was prescribed for the nodes near the angle of the mandible at the level of occlusal plane, improving the model mimicry and accuracy.[23]

Loading condition

This study was designed with the tripod occlusal contacts of the supporting cusps of premolar and both the molars loaded at buccal cusp tip, mesial, and distal fossa.[24] A load of 300 N[25] was applied as 37.5 N at each location in the vertical direction. The maximum equivalent von Mises stress values at the implant, cortical and trabecular bone and abutment tooth were recorded [Figure 4].
Figure 4: Application of load

Click here to view



   Results Top


The results obtained were tabulated [Table 2] and also shown in [Figure 5], [Figure 6], [Figure 7].
Table 2: Stress values around the cortical and trabecular bone, implant, and tooth after application of load

Click here to view
Figure 5: Stress value of Model 1 after application of vertical force

Click here to view
Figure 6: Stress value of Model 2 after application of vertical force

Click here to view
Figure 7: Stress value of Model 3 after application of vertical force

Click here to view


Stress at cortical bone

The maximum stress distribution in the cortical bone near the tooth was found in Model 2 (50.63 MPa) among the 3 models and less in Model 1 (35 MPa) and stress of cortical bone near the implant, the value was high in Model 3 (30.9 MPa) and again less in Model 1 as represented in [Graph 1].



Stress at trabecular bone

The value was significantly high in Model 1 in both trabecular bone near tooth and implant (6.15 MPa and 17.85 MPa) as compared to the other models and shown in [Graph 2].



Stress at implant

Less stress distribution was found in Model 1 (40.59 MPa) as compared to the other two. Maximum stress was transmitted in both Model 2 and 3 and the values were almost were same (44.22 MPa and 44.33 MPa, respectively) as shown in [Graph 3].



Stress at abutment tooth

The maximum stress value was lower around the abutment tooth in Model 2 (37 Mpa) and it was highest in Model 1 (59 MPa) as given in [Graph 4].



In Model 1 (with NRC placed on the mesial side of the implant), The values were 59.09 and 35.4 around the tooth and in the cortical bone region around the tooth, respectively. The highest stress value in the cortical bone around the implant was 20.43 and 40.59 on the mesiocervical region of the implant.

In Model 2 (with NRC placed on the distal side of the premolar tooth), the highest Von Mises stress values located in the cortical bone region around the tooth was 50.63 and 37 around the tooth. The highest stresses in the cortical bone region around the implant were 28.05 and 37 on the mesiocervical region of the implant.

In Model 3 (with NRC as occlusal rest on the distal surface of the pontic positioned on the rest seat prepared on the mesial side of the implant restoration), the highest stress values obtained were located on the cortical bone region around the tooth than around the tooth with the values 43.93 and 41.2, respectively. The highest stress values were generated around the implant than in the cortical bone with the values 44.3 and 30.94, respectively. In all the 3 models, the stress around the natural tooth and implant abutment was found to be gradually decreasing from the crestal region of the bone to the apical region.


   Discussion Top


In modern implantology, freestanding implants are the most preferred ones.[2],[26] A single implant can also serve as an abutment for several types of restoration including cantilever or tooth implant restoration.[8],[13],[27],[28] The use of cantilevered pontic supported by natural tooth as abutment in posterior areas has to be used cautiously because of the magnitude and axial direction of occlusal forces that are placed on the pontic of restoration.[16],[29],[30] The dissimilar mobility between an osseointegrated implant and tooth causes a dilemma in connecting an implant and tooth. The capability of a tooth to move and biomechanical aspects of tooth implant-supported prosthesis are significantly influenced by the viscoelastic properties of periodontal ligament. It plays a key role in long-term performance of tooth implant-connected prosthesis.[4],[5] A nonrigid connector is advised for homogeneous load distribution. It also relieves the stress and overload between the structures or long-span prosthesis.[31] Major risks that arise by connecting tooth and implant are bone loss around implant and natural tooth intrusion. The causes for tooth intrusion are multifactorial such as friction between the matrix and patrix wall in nonrigid connector and microjamming of food particles at the bottom of the matrix preventing tooth from reconnecting to its original position.

Chee and Jivraj reported a design to eliminate intrusion of tooth by using a modified nonrigid connector. A nonrigid connection in the form of a deep removable partial denture-type spoon-shaped rest was fabricated on the distal of pontic, and corresponding rest seat was prepared on the implant.[16] In this study, this was used to compensate the dissimilar mobility of tooth implant in tooth implant-supported prosthesis. The stress transmitted to the bone, implant, and tooth were evaluated after application of vertical load. The highest Von Mises stress values were taken into consideration and survival of the implant in all models was compared.

Duyck et al. demonstrated a better distribution of bending moments when metal was used as a prosthesis material in cantilevered or long-span prosthesis.[32] Ni-Cr alloy was used as a substructure in the study. Cibirka et al. compared the forces transmitted to human bone by gold, porcelain, and resin occlusal surfaces and found no statistically significant difference in the force absorption quotient of the occlusal surfaces.[33] In the current study, porcelain was used for occlusal surface.

Less Von Mises stress around implant was observed in Model 1 among the 3 models. This was due to the location of nonrigid connector between the implant and the pontic which allowed the implant to receive less stress, and greater part of the occlusal force was borne by the tooth. Ozcelik, Ersoy evaluated the stress around the implant and the abutment tooth with different location of nonrigid connector using 2D-FEM models and Photoelastic stress analysis method. There was a decrease in stress formed on the implant side in both the methods and in which nonrigid connector placed between implant and pontic.[22] A similar outcome was observed by Koosha and Mirhashemi in their study.[34] Beehelli suggested that nonrigid connector in tooth implant-supported prosthesis should be placed on the implant side to protect it from the torque effects.[35] He also indicated the advantages of this design such as allowing physical movement of the natural tooth, equal distribution of forces on the implant, and the natural tooth.

More stress distribution was found around the implant in Model 2. Kumar et al. noticed higher amount of stress around the implant than the tooth in a model where nonrigid connector was placed on the tooth side in their study.[36] The reason stated for this may be that additional movement of the pontic under occlusal forces caused the bridge to produce a cantilever effect on the implant. Thus, nonrigid connector placed on the tooth side had more disadvantages than the fixed prosthesis that is rigidly connected. The studies conducted by Misch and Ismail[37] and Breeding et al.[38] reported that there was no difference between the tooth implant-supported prosthesis with rigid connector and nonrigid connector placed on mesial side of the pontic.

No significant difference in stress value was found between Model 2 and 3. The nonrigid connection in a form of spoon-shaped occlusal rest could allow only a minimal movement and behave more or less like a rigid connection. A rigid restorative condition resolves more stress internally before it can reach the supporting tissues. This could be the reason for increased stress around implants and less stress around the natural tooth. Furthermore, in vivo and in vitro studies are needed to evaluate the biomechanical behavior of this kind of nonrigid connection.

More stress was found on the mesiocervical region of the implant in all the models.The reason could be disparity in the mobility of tooth and that of implant which was assumed as 100% osseointegrated. The bending moment which occurs after the applied static vertical load intrudes the natural tooth into the alveolus causes more stresses on the mesiocervical region of implant because the rotation center of implant is much higher than the natural tooth at the crestal bone level. Therefore, stress accumulation occurs in the cortical bone area due to the movement of the implant around this rotation.[22],[34] The higher amount of stress was noted around the natural tooth in Model 1 as compared to Model 2 because of the location of the nonrigid connector which allows receiving less stress.

All static masticatory forces applied were loaded vertically in this study. However, the masticatory forces are dynamic and may be directed oblique to the occlusal surface of tooth implant-supported prosthesis. Consequently, it is usually impossible to reproduce all details of natural behavior in a mathematical model which might influence the study outcomes.[39] Due to these limitations, the values obtained in this study may not resemble actual values but can definitely show the stress differences and advantages of various tooth implant-supported designs.


   Conclusions Top


The following conclusions may be drawn from the study:

  1. The nonrigid connector in tooth implant fixed prosthesis should be placed on the implant side to protect the implant from the torque effects
  2. There was no significant difference in the stress values around the implant between the nonrigid connector placed between abutment tooth and pontic and RPD type spoon-shaped rest extending from the distal surface of the pontic seats into rest preparation on the restored implant.


Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
   References Top

1.
Srinivasan M, Padmanabhan TV. Intrusion in implant-tooth-supported fixed prosthesis: An in vitro photoelastic stress analysis. Indian J Dent Res 2008;19:6-11.  Back to cited text no. 1
[PUBMED]  [Full text]  
2.
Shenoy VK, Rodrigue SJ, Prashanti E, Saldanha SJ. Tooth implant supported prosthesis: A literature review. J Interdiscip Dent 2013;3:143-50.  Back to cited text no. 2
    
3.
Al-Omiri MK, Al-Masri M, Alhijawi MM, Lynch E. Combined implant and tooth support: An up-to-date comprehensive overview. Int J Dent 2017;2017:6024565.  Back to cited text no. 3
    
4.
Pesun IJ. Intrusion of teeth in the combination implant-to-natural-tooth fixed partial denture: A review of the theories. J Prosthodont 1997;6:268-77.  Back to cited text no. 4
    
5.
Pesun IJ, Steflik DE, Parr GR, Hanes PJ. Histologic evaluation of the periodontium of abutment teeth in combination implant/tooth fixed partial denture. Int J Oral Maxillofac Implants 1999;14:342-50.  Back to cited text no. 5
    
6.
Akça K, Uysal S, Cehreli MC. Implant-tooth-supported fixed partial prostheses: Correlations between in vivo occlusal bite forces and marginal bone reactions. Clin Oral Implants Res 2006;17:331-6.  Back to cited text no. 6
    
7.
Kronström M, Trulsson M, Söderfeldt B. Patient evaluation of treatment with fixed prostheses supported by implants or a combination of teeth and implants. J Prosthodont 2004;13:160-5.  Back to cited text no. 7
    
8.
Pjetursson BE, Tan K, Lang NP, Brägger U, Egger M, Zwahlen M, et al. Asystematic review of the survival and complication rates of fixed partial dentures (FPDs) after an observation period of at least 5 years. Clin Oral Implants Res 2004;15:625-42.  Back to cited text no. 8
    
9.
Schlumberger TL, Bowley JF, Maze GI. Intrusion phenomenon in combination tooth-implant restorations: A review of the literature. J Prosthet Dent 1998;80:199-203.  Back to cited text no. 9
    
10.
Cordaro L, Ercoli C, Rossini C, Torsello F, Feng C. Retrospective evaluation of complete-arch fixed partial dentures connecting teeth and implant abutments in patients with normal and reduced periodontal support. J Prosthet Dent 2005;94:313-20.  Back to cited text no. 10
    
11.
Cohen SR, Orenstein JH. The use of attachments in combination implant and natural-tooth fixed partial dentures: A technical report. Int J Oral Maxillofac Implants 1994;9:230-4.  Back to cited text no. 11
    
12.
Lill W, Matejka M, Rambousek K, Watzek G. The ability of currently available stress-breaking elements for osseointegrated implants to imitate natural tooth mobility. Int J Oral Maxillofac Implants 1988;3:281-6.  Back to cited text no. 12
    
13.
Lin CL, Wang JC, Kuo YC. Numerical simulation on the biomechanical interactions of tooth/implant-supported system under various occlusal forces with rigid/non-rigid connections. J Biomech 2006;39:453-63.  Back to cited text no. 13
    
14.
Astrand P, Borg K, Gunne J, Olsson M. Combination of natural teeth and osseointegrated implants as prosthesis abutments: A 2-year longitudinal study. Int J Oral Maxillofac Implants 1991;6:305-12.  Back to cited text no. 14
    
15.
Oruc S, Eraslan O, Tukay HA, Atay A. Stress analysis of effects of nonrigid connectors on fixed partial dentures with pier abutments. J Prosthet Dent 2008;99:185-92.  Back to cited text no. 15
    
16.
Chee W, Jivraj S. Connecting implants to teeth. Br Dent J 2006;201:629-32.  Back to cited text no. 16
    
17.
Chee WW, Mordohai N. Tooth-to-implant connection: A systematic review of the literature and a case report utilizing a new connection design. Clin Implant Dent Relat Res 2010;12:122-33.  Back to cited text no. 17
    
18.
Wheeler's RC. Physiology and Occlusion. 8th ed. Philadelphia, PA: Saunders; 1974.  Back to cited text no. 18
    
19.
Shillingburg HT, Hobo S, Whitsett LD, Jacobi R, Brackett SE. Fundamentals of Fixed Prosthodontics. 3rd ed. Chicago: Quintessence; 1997.  Back to cited text no. 19
    
20.
Jenkins G. Precision Attachments – A Link to Successful Restorative Treatment. London: Quintessence Publication Co.; 1999.  Back to cited text no. 20
    
21.
Igor J Pesun. McCracken's Removable Partial Prosthodontics. 11th ed. St. Louis: Missouri; 2005.  Back to cited text no. 21
    
22.
Ozçelik T, Ersoy AE. An investigation of tooth/implant-supported fixed prosthesis designs with two different stress analysis methods: An in vitro study. J Prosthodont 2007;16:107-16.  Back to cited text no. 22
    
23.
Geng JP, Tan KB, Liu GR. Application of finite element analysis in implant dentistry: A review of the literature. J Prosthet Dent 2001;85:585-98.  Back to cited text no. 23
    
24.
Eskitascioglu G, Usumez A, Sevimay M, Soykan E, Unsal E. The influence of occlusal loading location on stresses transferred to implant-supported prostheses and supporting bone: A three-dimensional finite element study. J Prosthet Dent 2004;91:144-50.  Back to cited text no. 24
    
25.
Gibbs CH, Mahan PE, Lundeen HC, Brehnan K, Walsh EK, Sinkewiz SL, et al. Occlusal forces during chewing – Influences of biting strength and food consistency. J Prosthet Dent 1981;46:561-7.  Back to cited text no. 25
    
26.
Becker CM, Kaiser DA, Jones JD. Guidelines for splinting implants. J Prosthet Dent 2000;84:210-4.  Back to cited text no. 26
    
27.
Al Ansari B. Use of non rigid connection between natural teeth and implants to support fixed partial denture: Two years clinical evaluation. Saudi Dent J 1996;8:96-9.  Back to cited text no. 27
    
28.
Clarke DF, Chen ST, Dickinson AJ. The use of a dental implant as an abutment in three unit implant-tooth supported fixed partial denture: A case report and 32 month follow-up. Aust Dent J 2006;51:263-7.  Back to cited text no. 28
    
29.
Taylor TD, Agar JR, Vogiatzi T. Implant prosthodontics: Current perspective and future directions. Int J Oral Maxillofac Implants 2000;15:66-75.  Back to cited text no. 29
    
30.
Ormianer Z, Brosh T, Laufer BZ, Shifman A. Strains recorded in a combined tooth-implant restoration: An in vivo study. Implant Dent 2005;14:58-62.  Back to cited text no. 30
    
31.
Hita-Carrillo C, Hernández-Aliaga M, Calvo-Guirado JL. Tooth-implant connection: A bibliographic review. Med Oral Patol Oral Cir Bucal 2010;15:e387-94.  Back to cited text no. 31
    
32.
Duyck J, Van Oosterwyck H, Vander Sloten J, De Cooman M, Puers R, Naert I, et al. Influence of prosthesis material on the loading of implants that support a fixed partial prosthesis:In vivo study. Clin Implant Dent Relat Res 2000;2:100-9.  Back to cited text no. 32
    
33.
Cibirka RM, Razzoog ME, Lang BR, Stohler CS. Determining the force absorption quotient for restorative materials used in implant occlusal surfaces. J Prosthet Dent 1992;67:361-4.  Back to cited text no. 33
    
34.
Koosha S, Mirhashemi FS. An investigation of three types of tooth implant supported fixed prosthesis designs with 3D finite element analysis. J Dent (Tehran) 2013;10:51-63.  Back to cited text no. 34
    
35.
Bechelli AH. The osteointegrated prosthesis – Combination of osteointegrated implants and natural teeth in fixed prostheses. J Oral Implantol 1992;18:62-5.  Back to cited text no. 35
    
36.
Kumar GA, Kovoor LC, Oomen VM. Three dimensional finite element analysis of the stress distribution around the implant and tooth in tooth implant-supported fixed prosthesis design. J Dent Implant 2011;1:75-9.  Back to cited text no. 36
  [Full text]  
37.
Misch CM, Ismail YH. Finite element stress analysis of tooth-to-implant fixed partial denture designs. J Prosthodont 1993;2:83-92.  Back to cited text no. 37
    
38.
Breeding LC, Dixon DL, Sadler JP, McKay ML. Mechanical considerations for the implant tooth-supported fixed partial denture. J Prosthet Dent 1995;74:487-92.  Back to cited text no. 38
    
39.
Harutyunyan A, Pissiotis A. Biomechanical outcomes of tooth-implant-supported fixed prosthesis (FPP's) in periodontally healthy patients using root shape implants. Balk J Dent Med 2017;21:1-11.  Back to cited text no. 39
    

Top
Correspondence Address:
Dr. K Sumathi Nitin
Department of Prosthodontics and Crown and Bridge, Hitkarini Dental College and Hospital, Jabalpur, Madhya Pradesh
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/ijdr.IJDR_538_16

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6], [Figure 7]
 
 
    Tables

  [Table 1], [Table 2]



 

Top
 
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


    Abstract
   Introduction
    Materials and Me...
   Results
   Discussion
   Conclusions
    References
    Article Figures
    Article Tables