SHORT COMMUNICATION  
Year : 2014  |  Volume : 25  |  Issue : 4  |  Page : 539-540
Persistent radiographic cone cuts: A simple test to avoid the frustrating problem


1 Department of Oral Medicine and Radiology, Al-Badar Rural Dental College, Gulbarga, India
2 Department of Prosthodontics SDM College of Dental Sciences, Dharwad, Karnataka, India

Click here for correspondence address and email

Date of Submission01-Feb-2014
Date of Decision10-Feb-2014
Date of Acceptance31-May-2014
Date of Web Publication10-Oct-2014
 

   Abstract 

In routine dental radiography, one may encounter numerous radiographic errors, one of which may be partial images or cone cuts. These cones cut errors although may appear as simple problems, but often persistent cone cuts may be frustrating to the dental practitioner. In this study, our main aim was to find and solve the origin of the problem. This study was conducted with two different intraoral X-ray machines with 12 inch length position indicating devices (PIDs) with four No. 2 intraoral films. If the problem is originated in the X-ray machine, it can be solved by either tapping the collimator or by repositioning the PID.

Keywords: Collimator, cone cut, position indicating device

How to cite this article:
Shivanandappa SG, Mushannavar LS, Katti G. Persistent radiographic cone cuts: A simple test to avoid the frustrating problem. Indian J Dent Res 2014;25:539-40

How to cite this URL:
Shivanandappa SG, Mushannavar LS, Katti G. Persistent radiographic cone cuts: A simple test to avoid the frustrating problem. Indian J Dent Res [serial online] 2014 [cited 2019 Sep 20];25:539-40. Available from: http://www.ijdr.in/text.asp?2014/25/4/539/142572
Cone cut or partial images are often frustrating to the dental practitioner. Cone cut may be caused by various reasons. The most common being the image receptor is not in line with the position indicating device (PID)/radiographic cone. [1] Hence, the collimator which reduces the size and shape or cross-sectional area of the X-ray beam doesn't completely exposes the film resulting in cone cut. The other most common reason being the top of the film not completely immersed in the processing solutions. However, sometimes even after taking care of the above two mentioned facts; one may get these partial images/cone cuts persistently resulting in some frustration as well as embarrassment.


   Methods Top


In such cases, it is advisable to check for the X-ray machine itself for the correct alignment of the collimator with that of the PID or X-ray source. [2],[3] Hence, here we made such an attempt by conducting a simple experiment with two different X-ray machines. Four No. 2 dental films were placed on a piece of paper in the form of a cross-shaped template. The PID and the location of each film were traced on the paper [Figure 1], and the films were identified with small puncture holes in the film packet made with a pin, so each can be returned to its proper place after exposure [Figure 2]. The PID is positioned to cover approximately half of each film packet and to touch the packets [Figure 3]. The films were exposed using half the maximum anterior exposure, processed, and were returned to the proper order.
Figure 1: The position indicating device and the location of each film were traced on the paper

Click here to view
Figure 2: Films were identified with small puncture holes in the film packet made with a pin

Click here to view
Figure 3: The position indicating device is positioned to cover approximately half of each film packet and to touch the packets

Click here to view



   Discussion Top


We checked for the alignment of the image with that of the marked line of PID. If both were coinciding, then it was interpreted as collimator, X-ray source and PID are in line with each other as shown in [Figure 4]. If the image is not coinciding with that of the marked line as in [Figure 5], later it was interpreted as collimator, X-ray source and PID are in not in line with each other and may be the cause of persistent unavoidable partial images/cone cuts. [4] In such case, adjusting the collimator by tapping it into place or replacing the PID solves the problem. Hence, the collimator was corrected with the help of X-ray mechanic by tapping it to the correct place and once again image was taken [Figure 6]. This image [Figure 6] is showing the corrected alignment of collimator, X-ray source and PID. Thus, the frustrating cone cut/partial images may be avoided. This simple experiment is also cost-effective as it solves the root cause of the problem rather than replacing the defective machine with a new X-ray machine
Figure 4: Marked outline is in line with the image

Click here to view
Figure 5: Marked outline is not in line with the image

Click here to view
Figure 6: Corrected collimator, position indicating device and X-ray source again in line with each other

Click here to view
.

 
   References Top

1.Bailoor DN, Nagesh KS. Errors in dental radiography. Fundamentals of Oral Medicine and Radiology. 1 st ed. New Delhi: Jaypee Publications; 2005. p. 309.  Back to cited text no. 1
    
2.White SC, Pharoah MJ. Radiographic quality assurance and infection control. Oral Radiology; Principles and Interpretation. 5 th ed. Philadelphia: Elsevier Publications;2006. p. 113-4.  Back to cited text no. 2
    
3.Karjodkar FR. Radiographic prescription, quality control and Infection control. Textbook of Dental and Maxillofacial Radiology. 1 st ed. New Delhi: Jaypee Publications; 2006. p. 253.  Back to cited text no. 3
    
4.Ghom AG. Quality control in dental radiography. Textbook of Oral Radiology. 1 st ed. Philadelphia: Elsevier Publications; 2008. p. 223.  Back to cited text no. 4
    

Top
Correspondence Address:
Santosh Gowdru Shivanandappa
Department of Oral Medicine and Radiology, Al-Badar Rural Dental College, Gulbarga
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/0970-9290.142572

Rights and Permissions


    Figures

  [Figure 1], [Figure 2], [Figure 3], [Figure 4], [Figure 5], [Figure 6]



 

Top
 
 
 
  Search
 
    Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
    Email Alert *
    Add to My List *
* Registration required (free)  
 


    Abstract
   Methods
   Discussion
    References
    Article Figures

 Article Access Statistics
    Viewed4365    
    Printed32    
    Emailed1    
    PDF Downloaded148    
    Comments [Add]    

Recommend this journal